O2k signals and output: Difference between revisions

From Bioblast
No edit summary
No edit summary
Line 1: Line 1:
== Signal of the O2k-Core and add-on modules ==
== Signals of the O2k-Core and add-on modules ==
[[Image:Oxygraph-2k-Concept.jpg|right|300px|O2k-Concept.jpg]]
[[Image:Oxygraph-2k-Concept.jpg|right|300px|O2k-Concept.jpg]]
Three electronic channel types are available in the [[O2k-MultiSensor |O2k-MultiSensor system]]. All channels are available twofold (dual-data), for O2k-Chambers A and B (left and right), based on signals (numerical) sent at a fixed data sampling time interval (typical: 1 s or 2 s; range 0.2 s to >10 s):
Three electronic channel types are available in the [[O2k-MultiSensor |O2k-MultiSensor system]]. All channels are available twofold (dual-data), for O2k-Chambers A and B (left and right), based on signals (numerical) sent at a fixed data sampling time interval (typical: 1 s or 2 s; range 0.2 s to >10 s):
Line 6: Line 6:
: 2. Amp channel: The amperometric channel of the O2k (Amp plug), has electric current (ampere [Amp]) as the primary signal, which is converted to a voltage (volt [V]) as the raw signal. The Amp-Channel is used with the [[O2k-Fluo LED2-Module]] and the [[O2k-NO Amp-Module]].
: 2. Amp channel: The amperometric channel of the O2k (Amp plug), has electric current (ampere [Amp]) as the primary signal, which is converted to a voltage (volt [V]) as the raw signal. The Amp-Channel is used with the [[O2k-Fluo LED2-Module]] and the [[O2k-NO Amp-Module]].
: 3. pX channel with electric potential (volt [V]) as the primary and raw signal (BNC pX plug for the ion sensitive electrode, ISE; and separate plug for the pX reference electrode, Ref).
: 3. pX channel with electric potential (volt [V]) as the primary and raw signal (BNC pX plug for the ion sensitive electrode, ISE; and separate plug for the pX reference electrode, Ref).
: 4. Events: Automatic events (e.g. from the [[TIP2k]]) and manually triggered events carry information on a time point during the experiment, a short event name (text), and a longer string of comments (text).
 
<br />
'''System channels''', for instrumental control, not separated into O2k-Chamber A and B (single-data), based on signals (numerical) sent at the same fixed data sampling time interval:
'''System channels''', for instrumental control, not separated into O2k-Chamber A and B (single-data), based on signals (numerical) sent at the same fixed data sampling time interval:


: 5a. Block temperature: Measured block temperature equivalent to the experimental temperature in the O2k-Chambers at steady state [°C].
: 4a. Block temperature: Measured block temperature equivalent to the experimental temperature in the O2k-Chambers at steady state [°C].
: 5b. Peltier power (system channel): Relative heating or cooling power of the Peltier system.
: 4b. Peltier power (system channel): Relative heating or cooling power of the Peltier system.
: 5c. Barometric pressure (system channel): Measured barometric pressure [kPa].
: 4c. Barometric pressure (system channel): Measured barometric pressure [kPa].
: 5d. Ambient temperature (system channel): Measured temperature of room air (O2k-Series G upwards) [°C].
: 4d. Ambient temperature (system channel): Measured temperature of room air (O2k-Series G upwards) [°C].
: 5d. External temperature (system channel): Temperature measured with an independently placed thermistor (O2k-Series G upwards) [°C].
: 4d. External temperature (system channel): Temperature measured with an independently placed thermistor (O2k-Series G upwards) [°C].
 
'''Events''' carry information on a time point during the experiment, a short event name (text), and a longer string of comments (text):
: 5. Events: Automatic events (e.g. from the [[TIP2k]]) and manually triggered events.
<br />


'''NextGen-O2k''': additional channel types are available in the [[NextGen-O2k]]:
'''NextGen-O2k''': additional channel types are available in the [[NextGen-O2k]]:
: 6. Dual-fluo channel.
: 6. Dual-fluo channel.
: 7. Spectrophotometric channel.
: 7. Spectrophotometric channel.
Line 27: Line 31:
: '''II''': Flow, flux, rate. Examples: O<sub>2</sub> consumption; H<sub>2</sub>O<sub>2</sub> production; volume-specific flux is the negative or positive time derivative of concentration over time, frequently multiplied by 1,000 [pmol·s<sup>-1</sup>·ml<sup>-1</sup>]. Volume-specific flux normalized for the cell density [10<sup>6</sup>·ml<sup>-1</sup>] is flow [pmol·s<sup>-1</sup>·ml<sup>-1</sup>]. Volume-specific flux normalized for the mass density of the sample in the O2k-Chamber [mg·ml<sup>-1</sup>] is mass-specific flux [pmol·s<sup>-1</sup>·mg<sup>-1</sup>]. Flux in any respiratory state normalized for flux in a defined reference state is a rate [dimensionless] or [[flux control ratio]].
: '''II''': Flow, flux, rate. Examples: O<sub>2</sub> consumption; H<sub>2</sub>O<sub>2</sub> production; volume-specific flux is the negative or positive time derivative of concentration over time, frequently multiplied by 1,000 [pmol·s<sup>-1</sup>·ml<sup>-1</sup>]. Volume-specific flux normalized for the cell density [10<sup>6</sup>·ml<sup>-1</sup>] is flow [pmol·s<sup>-1</sup>·ml<sup>-1</sup>]. Volume-specific flux normalized for the mass density of the sample in the O2k-Chamber [mg·ml<sup>-1</sup>] is mass-specific flux [pmol·s<sup>-1</sup>·mg<sup>-1</sup>]. Flux in any respiratory state normalized for flux in a defined reference state is a rate [dimensionless] or [[flux control ratio]].
: '''III''': Force, potential (e.g. TMRM and Safranin for mt-membrane potential; redox potential from cytochrome or NADH redox states).
: '''III''': Force, potential (e.g. TMRM and Safranin for mt-membrane potential; redox potential from cytochrome or NADH redox states).
<br />
: '''IV''': Pre-calibrated signals for system channels.
<br />
: '''V''': Events.
<br />
<br />


Line 38: Line 46:
: 6. Single-fluo channel - [[NextGen-O2k]]: Plots of one selected wavelength can be selected for recording. Output type I(1), II(1), III(1) depends on the fluorophore.
: 6. Single-fluo channel - [[NextGen-O2k]]: Plots of one selected wavelength can be selected for recording. Output type I(1), II(1), III(1) depends on the fluorophore.
: 7. Dual-fluo channel - [[NextGen-O2k]]: Plots of two selected wavelengths can be selected for simultaneous recording. Output type I(2), II(2), III(2) depends on the fluorophore.
: 7. Dual-fluo channel - [[NextGen-O2k]]: Plots of two selected wavelengths can be selected for simultaneous recording. Output type I(2), II(2), III(2) depends on the fluorophore.
: 8. Spectrophotometric channel - [[NextGen-O2k]]: Output type IV (three plots for redox states of cytochromes ''aa''<sub>3</sub>, ''b'', and ''c''.
: 8. Spectrophotometric channel - [[NextGen-O2k]]: Output type VI (three plots for redox states of cytochromes ''aa''<sub>3</sub>, ''b'', and ''c''.

Revision as of 17:11, 28 July 2015

Signals of the O2k-Core and add-on modules

O2k-Concept.jpg

Three electronic channel types are available in the O2k-MultiSensor system. All channels are available twofold (dual-data), for O2k-Chambers A and B (left and right), based on signals (numerical) sent at a fixed data sampling time interval (typical: 1 s or 2 s; range 0.2 s to >10 s):

1. O2 channel with electric current (ampere [Amp]) as the primary signal, which is converted to a voltage (volt [V]) as the raw signal. The O2k-Core supports not only the O2 channel (O2 plug), but also the channel types 2 and 3 for the add-on modules.
2. Amp channel: The amperometric channel of the O2k (Amp plug), has electric current (ampere [Amp]) as the primary signal, which is converted to a voltage (volt [V]) as the raw signal. The Amp-Channel is used with the O2k-Fluo LED2-Module and the O2k-NO Amp-Module.
3. pX channel with electric potential (volt [V]) as the primary and raw signal (BNC pX plug for the ion sensitive electrode, ISE; and separate plug for the pX reference electrode, Ref).

System channels, for instrumental control, not separated into O2k-Chamber A and B (single-data), based on signals (numerical) sent at the same fixed data sampling time interval:

4a. Block temperature: Measured block temperature equivalent to the experimental temperature in the O2k-Chambers at steady state [°C].
4b. Peltier power (system channel): Relative heating or cooling power of the Peltier system.
4c. Barometric pressure (system channel): Measured barometric pressure [kPa].
4d. Ambient temperature (system channel): Measured temperature of room air (O2k-Series G upwards) [°C].
4d. External temperature (system channel): Temperature measured with an independently placed thermistor (O2k-Series G upwards) [°C].

Events carry information on a time point during the experiment, a short event name (text), and a longer string of comments (text):

5. Events: Automatic events (e.g. from the TIP2k) and manually triggered events.


NextGen-O2k: additional channel types are available in the NextGen-O2k:

6. Dual-fluo channel.
7. Spectrophotometric channel.


O2k output

The 'output' of a channel is the finally important infomation, which may be summarized in a data base. Three types of output can be derived from the signal of the O2k channels:

I: Concentration. Examples: O2; Ca2+; amount of substrance per volume [1 µM = 1 µmol·dm-3 = 1 nmol·ml-1].
II: Flow, flux, rate. Examples: O2 consumption; H2O2 production; volume-specific flux is the negative or positive time derivative of concentration over time, frequently multiplied by 1,000 [pmol·s-1·ml-1]. Volume-specific flux normalized for the cell density [106·ml-1] is flow [pmol·s-1·ml-1]. Volume-specific flux normalized for the mass density of the sample in the O2k-Chamber [mg·ml-1] is mass-specific flux [pmol·s-1·mg-1]. Flux in any respiratory state normalized for flux in a defined reference state is a rate [dimensionless] or flux control ratio.
III: Force, potential (e.g. TMRM and Safranin for mt-membrane potential; redox potential from cytochrome or NADH redox states).


IV: Pre-calibrated signals for system channels.


V: Events.


1. O2 channel - O2k-Core: Output type I and II.
2a. Amp channel - O2k-Fluo LED2-Module: The focus of the output depends on the specific fluorophore. Output type I, II, or III, or any combination may relate to the primary level of interest.
2b. Amp channel - O2k-NO Amp-Module: Output type I and II.
3a. pX channel - O2k-TPP+ ISE-Module: Output type III.
3b. pX channel - O2k-pH ISE-Module: Output type I and II.
4. Events - O2k-Core and TIP2k
6. Single-fluo channel - NextGen-O2k: Plots of one selected wavelength can be selected for recording. Output type I(1), II(1), III(1) depends on the fluorophore.
7. Dual-fluo channel - NextGen-O2k: Plots of two selected wavelengths can be selected for simultaneous recording. Output type I(2), II(2), III(2) depends on the fluorophore.
8. Spectrophotometric channel - NextGen-O2k: Output type VI (three plots for redox states of cytochromes aa3, b, and c.
Cookies help us deliver our services. By using our services, you agree to our use of cookies.