== Technical service == | |||
'''Volume calibration of the [[O2k-Chamber]]''' | |||
* Loosen the stopper fixation rings (A) and (B) with the L-shaped Allen key (OroboPOS-Service Kit). | |||
* Follow the instructions in the O2k-Manual: [[MiPNet19.18A O2k-Start]] | |||
* Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring. | |||
Β | |||
Β | |||
{{#set:Technical service=Chamber}} | |||
Β | |||
[[Category:Technical service]] | |||
Β | |||
__SHOWFACTBOX__ |
Revision as of 08:40, 9 May 2015
Volume-Calibration Ring
Description | Volume-Calibration Ring (A or B) for PVDF or PEEK O2k-Stoppers, white; attached to the stoppers. |
---|---|
Product ID | 2432A-01 or 2432B-01 |
Type | O2k, Stopper |
Link | MiPNet12.06 |
Image |
Technical service
Volume calibration of the O2k-Chamber
- Loosen the stopper fixation rings (A) and (B) with the L-shaped Allen key (OroboPOS-Service Kit).
- Follow the instructions in the O2k-Manual: MiPNet19.18A O2k-Start
- Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.