== Very low temperatures ==
'''Question:''' Is it possible to use the oxygraph with a chamber temperature of 0 Β°C?
Our system is specified for a chamber temperature down to 4 Β°C at 25 Β°C room temperature and down to 2 Β°C at "lower ambient temperature", see
[http://www.oroboros.at/index.php?id=o2k-specifications| specifications].
Lower ambient temperature does not necessarily mean a "cool room". For a chamber temperature of 2 Β°C, an air conditioned room will usually suffice.
While we can only guarantee the working of the instrument within these limits, it is possible to reach a chamber temperature of 0 Β°C at a "lower ambient temperature". In general, however, thisΒ  may require a "cool room". For this purpose, the main unit itself can beΒ  exposed to ambient temperatures down to just above 0Β°C (all the electronicΒ  components are rated at least down to 0 Β°C, but 0 Β°C room temperature should be avoided due to condensation concerns). Also, for very low room temperatures above 0 Β°C, some means of humidity control should be used.
We recently performed a short test and were able to reach 0 Β°C chamber temperature quite easily at a room temperature of about 14 Β°C.Β  In order to maintain this temperature, the cooling power of the Peltier element reached 95% of maximum. Thus, a "cool room" would probably be necessary for longer experiments.
'''Update for O2k Series E:'''
Because Series E Oxygraphs have a more efficient temperature management system the experiment described above was repeated with a Series E O2k. In this case it was not necessary to reduce the room temperature. [[User:Fasching Mario|Fasching Mario]] 14:39, 4 February 2013 (CET)
Details: O2K - 0Β°C block temp.Β  - stable Peltier power was reached after about 1:05 h from setting blockΒ  temp. from stable 37Β°C to 0Β°C. - stable Peltier power was about -83 %. - room temperature was 21.1Β°C [[User:Capek Ondrej|Capek Ondrej]] 14:36, 4 February 2013 (CET)