Difference between revisions of "O2k-Main Unit"

From Bioblast
Β 
(18 intermediate revisions by 8 users not shown)
Line 1: Line 1:
{{Product
{{Technical support}}
__TOC__
{{MitoPedia without banner
|abbr=
|description=The '''O2k-Main Unit''' is a component of the [[O2k-Core]]. The O2k-Main Unit consists of functionally defined, integral elements, the ([[O2k-Main Basic]], [[O2k-Peltier Temperature Control]], two [[O2k-Electromagnetic Stirrer Twin-Control]] units, two [[O2k-Amperometric OroboPOS Twin-Channel]]s, [[O2k-Barometric Pressure Transducer]]), which cannot be obtained separately.
|description=The '''O2k-Main Unit''' is a component of the [[O2k-Core]]. The O2k-Main Unit consists of functionally defined, integral elements, the ([[O2k-Main Basic]], [[O2k-Peltier Temperature Control]], two [[O2k-Electromagnetic Stirrer Twin-Control]] units, two [[O2k-Amperometric OroboPOS Twin-Channel]]s, [[O2k-Barometric Pressure Transducer]]), which cannot be obtained separately.
|product id=20000
|info=[[Oroboros O2k]]
|product type=O2k-Core
|info=[[O2k-Core]]
}}
}}
__TOC__
The '''O2k-Main Unit''' consists of integral elements, which are not separately available:
The '''O2k-Main Unit''' consists of integral elements, which are not separately available:
{{#ask: mainlabel=Title|[[Category:Products]] | [[Product type::O2k-Main Unit]]
{{#ask: [[Category:Products]] [[Product type::O2k-Main Unit]]
| ?description
| ?description
| ?product id
| ?product id
Line 12: Line 17:
|sort=product id
|sort=product id
|order=ascending
|order=ascending
|mainlabel=Title
}}
}}


__TOC__
== O2k series and support of O2k-Modules ==
::::Β» [[O2k series]]




== O2k-Series ==
{{MitoPedia O2k and high-resolution respirometry
The O2k is continuously developed further to integrate various innovations. While all O2k instruments have the same high standard for measuring respiration, there are differences with respect to:
|mitopedia O2k and high-resolution respirometry=O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry
* Type of [[Catalogue O2k-Stoppers|stoppers]] and [[O2k-Chamber Holder|chamber holders]] used
}}
* Type of [[OroboPOS-Connector]]
* Availability of additional electronic channels for O2k-MultiSensor applications.
[[Image:Serial_number_sticker_small.JPG|right|200 px|alt=The serial number of each O2k is shown on a sticker at the rear of the O2k.]]
The '''Series''' of the O2k is specified as the capital letter in the serial number of the O2k, shown on the label on the rear of the O2k-Main Unit [[MiPNet12.06]].Β  A serial number B-xxxx or C-xxxx denotes O2k-Series B or C, while D-xxxx denotes an O2k-Series D.
Β 
Β 
Β 
Β 
[[Image:Serial_number_signal_line.png|right|200 px|alt=On-line the serial number is displayed in the signal line]]
With DatLab operating connected to the O2k, the serial number of the O2k is displayed:
Β 
* in the right corner of the status line, besides the DatLab version number (bottom).
[[Image:Serial_number_control_window.png|right|200 px|alt=On-line the serial number is displayed in the window caption of the O2k-Control window.]]
* in the window caption of the O2k-Control window [F7].
Β 
== Support of O2k-Modules ==
The O2k [[Electronic-TIP2k Upgrading\O2k-Main Unit Series E|(Series E)]] and hgher support all modules. O2k [[Electronic-TIP2k Upgrading\O2k-Main Unit Series A-D|(Series A-D)]] require the [[O2k-Catalogue: O2k-MultiSensor\Electronic Upgrading|O2k-MultiSensor electronic upgrade]]. An exception is the application of an NO sensor in a single chamber of the Series A to C O2k.
Β 
== Power supply cable ==
The O2k is delivered with one of three different types of power supply cables:Β  Β  Β  Β  Β 
* US type ([[O2k-Main Power Cable\120 V\US-CA]], 21112-01)
* European type ([[O2k-Main Power Cable\230 V\Europe]], 21111-01)
* Australian Type ([[O2k-Main Power Cable\230 V\AU-NZ]], 21113-01)
Β 
Use 21111-01 O2k-Main Power Cable\230 V\Europe (Hybrid German/French plug CEE 7-7. Type E+F) only for compatible socket. Note that some '''Danish''' sockets also accept the CEE7/7 plugs, but there is '''no grounding''' connection with these plugs because a male ground pin is required on the plug.
Β 
== Control lights ==
After switching on the O2k-Main Unit, the control LEDs on the O2k are illuminated.
Β 
* On the left hand side: the lights should be green.
* On the right hand side: the 'Comm' light should be yellow (before connection to DatLab) or green (after connection to DatLab during data recording).
* Left and right stirrer: green (stirrers are on and running at set speed); yellow (briefly the stirrers are accelerating after having been switched on); red (stirrers are switched off).
Β 
Should these lights be unlit, you should exchange the fuse on the rear of the O2k-Main Unit where the power cable is plugged in. A spare fuse is supplied with the O2k in the accessory box ([[O2k-Assembly Kit]]). If the 'Comm' light is red, this indicates an electronic problem.

Latest revision as of 12:05, 14 April 2021


                  


O2k-Open Support

O2k-Main Unit



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

The O2k-Main Unit is a component of the O2k-Core. The O2k-Main Unit consists of functionally defined, integral elements, the (O2k-Main Basic, O2k-Peltier Temperature Control, two O2k-Electromagnetic Stirrer Twin-Control units, two O2k-Amperometric OroboPOS Twin-Channels, O2k-Barometric Pressure Transducer), which cannot be obtained separately.


Reference: Oroboros O2k



The O2k-Main Unit consists of integral elements, which are not separately available:


O2k series and support of O2k-Modules

Β» O2k series


MitoPedia O2k and high-resolution respirometry: O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry