MiPNet09.12 O2k-Titrations: Difference between revisions
From Bioblast
Beno Marija (talk | contribs) No edit summary |
No edit summary |
||
Line 7: | Line 7: | ||
|abstract='''Oroboros (2018) O2k manual titrations: SUIT protocols with mitochondrial preparations. Mitochondr Physiol Network 09.12(17): 1-2.'''ย ย | |abstract='''Oroboros (2018) O2k manual titrations: SUIT protocols with mitochondrial preparations. Mitochondr Physiol Network 09.12(17): 1-2.'''ย ย | ||
A list of substrates, uncouplers, and inhibitors used in [[SUIT]] protocols in [[ | A list of substrates, uncouplers, and inhibitors used in [[SUIT]] protocols in [[high-resolution respirometry]], with information on abbreviations, stock concentrations, final experimental concentrations, volumes titrated into the 2 mL O2k-chamber, and syringes used for manual titrations. | ||
:ยป [[O2k-Protocols]] | :ยป [[O2k-Protocols]] | ||
:ยป Product: [[Oroboros O2k]], [[Oroboros O2k-Catalogue |O2k-Catalogue]] | :ยป Product: [[Oroboros O2k]], [[Oroboros O2k-Catalogue |O2k-Catalogue]] |
Revision as of 11:04, 17 March 2019
O2k manual titrations: SUIT protocols with mitochondrial preparations. |
ยป ยปVersions
Oroboros (2018-10-01) Mitochondr Physiol Network
Abstract: Oroboros (2018) O2k manual titrations: SUIT protocols with mitochondrial preparations. Mitochondr Physiol Network 09.12(17): 1-2.
A list of substrates, uncouplers, and inhibitors used in SUIT protocols in high-resolution respirometry, with information on abbreviations, stock concentrations, final experimental concentrations, volumes titrated into the 2 mL O2k-chamber, and syringes used for manual titrations.
- ยป O2k-Protocols
- ยป Product: Oroboros O2k, O2k-Catalogue
โข O2k-Network Lab: AT_Innsbruck_Oroboros
Labels: MiParea: Instruments;methods
Regulation: ADP, Inhibitor, Substrate, Uncoupler, Amino acid
HRR: O2k-Protocol
O2k-chemicals and media
Note: Titration of chemicals in respirometry
- Overshoots or short interruptions of the flux after the injection of chemicals do not have a meaning in respirometric measurements. A disturbance of the signal is only observed due to the injected chemical and does not represent mitochondrial respiration. With the addition of a chemical solved in ethanol a little bit of oxygen is always introduced, since ethanol has higher oxygen solubility than water.
Further information
- ยป MitoPedia: SUIT - Substrate-uncoupler-inhibitor-titration (SUIT) protocols for respirometry
- ยป MiPNet03.02 - Chemicals and media for respirometry
- ยป MiPNet19.14 - SOP for manual O2k-titrations with Hamilton microsyringes.