High signal at zero oxygen: Difference between revisions
From Bioblast
No edit summary |
Tindle Lisa (talk | contribs) mNo edit summary |
||
Line 9: | Line 9: | ||
:::[[File:High zero current test 2.png|1000px]] | :::[[File:High zero current test 2.png|1000px]] | ||
::::*Note: Before sending us the O2k or any spare components, please contact us via our O2k-Open Support system: [ | ::::*Note: Before sending us the O2k or any spare components, please contact us via our O2k-Open Support system: [https://www.oroboros.at/index.php/support/ O2k-Open Support] | ||
:::::References: ย | :::::References: ย | ||
::::::[1] [https://wiki.oroboros.at/index.php/OroboPOS-Connector_MitoPedia#Cleaning_the_electrical_connections Cleaning the electrical connections] | ::::::[1] [https://wiki.oroboros.at/index.php/OroboPOS-Connector_MitoPedia#Cleaning_the_electrical_connections Cleaning the electrical connections] |
Revision as of 18:06, 3 November 2020
High signal at zero oxygen |
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Description
'A high signal at zero oxygen' may be observed during zero calibration. The following instructions show how to further localize and possibly solve this technical problem.
Abbreviation: R0
Reference: Zero calibration, OroboPOS-Connector
High zero current
-
- Note: Before sending us the O2k or any spare components, please contact us via our O2k-Open Support system: O2k-Open Support
-
- Example: First, the OroboPOS-Connector and the POS were disassembled and disconnected from the main unit. Second, the OroboPOS-Connector was connected to the O2k-Main unit without the OroboPOS. Third, the OroboPOS was connected to the OroboPOS-Connector without membranes. After each step, the O2 raw signal [V] was recorded for 10 min. The marks were set to the O2 raw signal [V] to evaluate the zero current.
- Figure 1. High zero current test. Upper panel: Zero current is in the range of ยฑ0.00099 V. Lower panel: Zero current is higher than 0.00099 V when the OroboPOS-Connector is connected to the O2k-Main unit and when the OroboPOS is connected to the OroboPOS-Connector, which means that both the OroboPOS-Connector and OroboPOS might have electrical problems.
Keywords
- Bioblast links: Oxygen signal - >>>>>>> - Click on [Expand] or [Collapse] - >>>>>>>
- Specific
- ยป Oxygen signal
- ยป O2k signal line
- Specific
- O2k-Procedures
- MiPNet O2k-Procedures
- General
- ยป Oxygen, dioxygen, O2
- ยป Oxygen calibration - DatLab
- ยป Oxygen solubility
- ยป Oxygen solubility factor
- ยป Oxygen pressure
- ยป Concentration
- ยป Activity
- ยป Pressure - Pascal
- ยป Barometric pressure
- ยป High-resolution respirometry
- ยป OroboPOS
- ยป Polarographic oxygen sensor
- ยป MitoFit Quality Control System
- ยป Sensitivity
- General
- Other keyword lists
MitoPedia O2k and high-resolution respirometry: O2k hardware, O2k-Open Support, O2k-Respirometry