Volume-Calibration Ring: Difference between revisions
From Bioblast
No edit summary |
No edit summary |
||
Line 14: | Line 14: | ||
::::* Follow the instructions in the O2k-Manual: [[MiPNet22.11 O2k-FluoRespirometer manual]] | ::::* Follow the instructions in the O2k-Manual: [[MiPNet22.11 O2k-FluoRespirometer manual]] | ||
::::* Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring. | ::::* Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring. | ||
Revision as of 11:58, 27 April 2020
Volume-Calibration Ring
Description | Volume-Calibration Ring (A or B) PVDF for O2k-Stoppers, white; attached to the stoppers. |
---|---|
Product ID | 2432A-01 or 2432B-01 |
Type | O2k, Stopper |
Link | MiPNet22.11 O2k-FluoRespirometer manual |
Image |
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Technical support
- Volume calibration of the O2k-Chamber
- Loosen the stopper fixation rings (A) and (B) with the Allen key (OroboPOS-Service Kit).
- Follow the instructions in the O2k-Manual: MiPNet22.11 O2k-FluoRespirometer manual
- Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.
- Volume calibration of the O2k-Chamber