- Currently Answered Questions from the »Oroboros Ecosystem Agenda«
Feb 2024
- Feb-22 Th
What to do if the glass of the O2k-chamber starts to become opaque?
"Turbidity on the glass wall may be caused by precipitated protein. Remove the chambers from the O2k and with the O2k-stirrers removed, immerse in a beaker filled with HCl (10 M) overnight under a hood."
-»MiPNet19.03_O2k-cleaning_and_ISS«
- communicated by »Cardoso Luiza« and »Leo Elettra«
- Feb-08 Th
Can I use the same stock solutions for all reagents for different O2k-chamber volumes?
"For some reagents, the concentration of the stock solution differs between the 2.0-mL and the 0.5-mL chamber (O2k-sV-Module) to facilitate a precise volume titration."
-»MiPNet09.12 O2k-Titrations«
- communicated by »Timon-Gomez Alba« and »Leo Elettra«
Jan 2024
- Jan-25 Th
How do I view and change the settings of a measurement/file in DatLab 8?
"O2k control (active) will always show the settings of the active recording [and allow changes], while O2k info (view) will show the settings of the recorded selected file."
-»MiPNet28.07 NextGen-O2k Series XB manual«; »MiPNet28.08 Oroboros O2k Series J manual«
- communicated by »Baglivo Eleonora« and »Grings Mateus«
- Jan-11 Th
What are special features of the new SmartPOS (O2k-Series J and NextGen-O2k-Series XB)?
- "The electrical connections of the SmartPOS are protected thus there is no need to clean them."
- "The serial number of the SmartPOS is automatically recognized by the DatLab 8 software."
»MiPNet28.10 SmartPOS-service«
- communicated by »Grings Mateus« and »Baglivo Eleonora«
- Jan-08 Mo
Do you need to export DatLab traces for your publication?
Save your graph as an editable picture using "Copy to clipboard" in DatLab 7.4 (.wmf) or "Save" in DatLab 8 (.svg).
- »Copy to clipboard in DatLab 7«, »Download DatLab 8«
- communicated by »Baglivo Eleonora« and »Grings Mateus«