== May 2023 ==
*'''May- 09 Tu'''
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Yellow-take care.png|25px|Info: take care]]
|'''Does your polarographic oxygen sensor (POS) need a deep clean?'''<br>
A POS with a very dark anode might need several rounds of ammonia cleaning, possibly including prolonged exposure overnight to restore performance specifications.
<br>
- »'''[[MiPNet19.18B POS-service|POS service]]'''«<br>
- communicated by »[[Cecatto Cristiane]]« and »[[Leo Elettra]]«
|}
----
== Mar 2023 ==
== Mar 2023 ==
*'''Mar- 28 Tu'''
*'''Mar- 28 Tu'''

Revision as of 09:43, 30 May 2023

Currently Answered Questions from the »Oroboros Ecosystem Agenda«

May 2023

  • May- 09 Tu
O2k-Open SupportInfo: take care Does your polarographic oxygen sensor (POS) need a deep clean?

A POS with a very dark anode might need several rounds of ammonia cleaning, possibly including prolonged exposure overnight to restore performance specifications.
- »POS service«
- communicated by »Cecatto Cristiane« and »Leo Elettra«


Mar 2023

  • Mar- 28 Tu
O2k-Open SupportInfo: improve Unusual stirrer bar rotation?

A “jumping” stirrer or a slow rotation of the stirrer bar may indicate a problem either with the stirrer itself or with the cleaning of the glass chamber.
- »Stirrer bar - technical support«
- communicated by »Baglivo Eleonora« and »Grings Mateus«

  • Mar- 21 Tu
O2k-Open SupportWarning: loss of data Did you know that a humid environment and/or liquid spillage may lead to a high raw POS signal [V]?

See how to clean and dry the electrical connections of the OroboPOS-Connector.
- »Cleaning the electrical connections«
- communicated by »Cecatto Cristiane« and »Baglivo Eleonora«

Feb 2023

  • Feb- 23 Th
O2k-Open SupportInfo: improve Add catalase when measuring Complex IV (CIV) activity in permeabilized fibers (pfi)

Pfi have a high content of iron-containing proteins, which accelerate the autoxidation of tetramethyl-p-phenylenediamine (TMPD) in the CIV assay. Catalase prevents the accumulation of H2O2 generated by this autoxidation.
- »Chemical Background« and »Complex IV«
- communicated by »Cecatto Cristiane« and »Schmitt Sabine«


  • Feb- 16 Th
O2k-Open SupportInfo: improve Do you observe simultaneous fluctuations in the O2 slope neg. and the Peltier power?

This can indicate fluctuations in the room temperature or electrical supply. Controlling the temperature or changing the electrical supply may resolve this problem.
- »Troubleshooting«
- communicated by »Cecatto Cristiane« and »Schmitt Sabine«