Line 1: | Line 1: | ||
== Details == | == Details == | ||
In the [[O2k-Fluo LED2-Module]]dye based absorption filters in form of polymer films are used. Stating cut-off values or otherwise numerical describing the filters (short of a full spectrum) does make little sense. However, if required the following values can be used: | In the [[O2k-Fluo LED2-Module]] dye based absorption filters in form of polymer films are used. Stating cut-off values or otherwise numerical describing the filters (short of a full spectrum) does make little sense. However, if required the following values can be used: | ||
*Saf excitation filter: (SG357): dye based short pass filter, 50% T: 470 nm, suppressing (<5% Transmission) lambda > 530 nm | *Saf excitation filter: (SG357): dye based short pass filter, 50% T: 470 nm, suppressing (<5% Transmission) lambda > 530 nm | ||
*Saf emission filter: (SG19): dye based long pass filter, 50% T: 600 nm, suppressing (<1% transmission) lambda < 550 nm | *Saf emission filter: (SG19): dye based long pass filter, 50% T: 600 nm, suppressing (<1% transmission) lambda < 550 nm |
Latest revision as of 15:00, 1 December 2014
Details
In the O2k-Fluo LED2-Module dye based absorption filters in form of polymer films are used. Stating cut-off values or otherwise numerical describing the filters (short of a full spectrum) does make little sense. However, if required the following values can be used:
- Saf excitation filter: (SG357): dye based short pass filter, 50% T: 470 nm, suppressing (<5% Transmission) lambda > 530 nm
- Saf emission filter: (SG19): dye based long pass filter, 50% T: 600 nm, suppressing (<1% transmission) lambda < 550 nm