Volume-Calibration Ring sV

From Bioblast


                  


O2k-Open Support

Volume-Calibration Ring sV



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

Volume calibration ring A+B PVDF for Stopper.jpg

Volume-Calibration Ring sV (A or B), white PVDF, attached to PEEK O2k-Stopper sV.


Reference: Product details and purchase information



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Technical support

Volume calibration of the 0.5-mL O2k-chamber
  • Loosen the stopper fixation rings (A) and (B) with the Allen key.
  • Follow the instructions in the O2k-Manual: MiPNet22.11 O2k-FluoRespirometer manual
  • Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.
Volume calibration ring A+B aluminum for Stopper.jpg