Volume-Calibration Ring

From Bioblast
Jump to: navigation, search

Header ecosystem 2020.jpg

News - Agenda         O2k-Open Support         O2k-Publications         O2k-Workshops and Events         O2k-Network         O2k-Feedback         NextGen-O2k



Volume-Calibration Ring

O2k-Catalogue

Description Volume-Calibration Ring (A or B) PVDF for O2k-Stoppers, white; attached to the stoppers.
Product ID 2432A-01 or 2432B-01
Type O2k, Stopper
Link MiPNet22.11 O2k-FluoRespirometer manual
Image
Volume calibration ring A+B PVDF for Stopper.jpg

Template NextGen-O2k.jpg
O2k-Open Support


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Technical support

Volume calibration of the O2k-Chamber
  • Loosen the stopper fixation rings (A) and (B) with the Allen key (OroboPOS-Service Kit).
  • Follow the instructions in the O2k-Manual: MiPNet22.11 O2k-FluoRespirometer manual
  • Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.