Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Talk:Different O2 fluxes in left and right chamber

From Bioblast

Hardware defects: As hardware defects (sensors, electronics, etc.) have many negative effects (noisy signal, slow response, no signal, etc.) on experiments, it is difficult to see how a real hardware defect can cause a systematic error in flux calculation once a correct calibration of the O2 sensor (at air saturation and at zero oxygen) has been obtained. Fasching Mario 09:49, 7 November 2014 (CET)